Simulation and experimental study on the influence of cables on the performance of Search Coil magnetometers
-
摘要: 感应式磁传感器线圈与前置放大电路之间的电缆长度对信号传输有着显著的影响。本文对带有较长电缆的感应式磁传感器进行了建模仿真分析和试验验证。在以10Hz-1kHz频带宽度,30fT/√Hz@1kHz的噪声水平为目标的感应式磁强计样机基础上,研究并建立了感应线圈-电缆-前置放大电路的电路原理模型,分析了电缆在基于跨阻抗前置放大电路的感应式磁强计中的作用,并进行了理论仿真。通过试验验证了电缆长度变化对传感器灵敏度和噪声频率分布的影响,其结果与理论模型基本一致,说明电缆长度的增加对于感应式磁传感器的影响虽然显著,但是可以通过理论模型进行设计和预测。电缆长度增加会减小低频灵敏度,降低低频噪声,但是也会显著提高高频的噪声水平,对于低频段的磁波动探测影响不大,但是不利于高频段的磁场探测。对于目标传感器而言,电缆长度即使达到50米,10Hz~1kHz频段内仍然有较好的响应和较低的噪声水平,能够满足30fT/√Hz@1kHz的要求。
Abstract: The length of the cable between the search coil and the preamplifier circuit has a significant impact on signal transmission. This study conducts a multidisciplinary simulation analysis and experimental verification on search coil magnetometer with long-distance cables. Based on search coil magnetometer prototype targeting a frequency bandwidth of 10Hz-1kHz and a noise level of 30fT/√Hz@1kHz, this study conducts research and establishes a circuit-principle model of the search coil-cable-preamplifier circuit. The role of the cable in search coil magnetometer based on transimpedance preamplifier circuit is analyzed, and theoretical simulation is carried out. The influence of cable length variation on sensitivity and noise frequency distribution is verified through experiments, and the results are basically consistent with the theoretical model, indicating that although the increase in cable length has a significant impact on search coil magnetometer, it can be designed and predicted through theoretical models. An increase in cable length reduces low-frequency sensitivity and low-frequency noise level, but it also significantly raises the level of high-frequency noise. It has little effect on magnetic wave detection at low-frequency, but substantially degrades magnetic field detection performance at high-frequency. For the target search coil, even if the cable length reaches 50 meters, it demonstrates satisfactory response and low noise levels within the 10 Hz–1 kHz frequency band, meeting the requirement of 30fT/√Hz@1kHz.-
Key words:
- Space Exploration /
- Search Coil Magnetometer /
- Transimpedance Amplifier /
- Noise
-
-
计量
- 文章访问数: 23
- HTML全文浏览量: 0
- PDF下载量: 0
-
被引次数:
0(来源:Crossref)
0(来源:其他)