Volume 9 Issue 3
Aug.  1989
Turn off MathJax
Article Contents
Zhou Huai-bei, Wang Wen-qing, Sun Chuan-li. THE INFLUENCE OF HIGH ENERGY ELECTRONS AND CONTAMINATION ON LANGMUIR PROBE DIAGNO-STICS OF GAS DISCHARGED[J]. Chinese Journal of Space Science, 1989, 9(3): 180-186. doi: 10.11728/cjss1989.03.180
Citation: Zhou Huai-bei, Wang Wen-qing, Sun Chuan-li. THE INFLUENCE OF HIGH ENERGY ELECTRONS AND CONTAMINATION ON LANGMUIR PROBE DIAGNO-STICS OF GAS DISCHARGED[J]. Chinese Journal of Space Science, 1989, 9(3): 180-186. doi: 10.11728/cjss1989.03.180

THE INFLUENCE OF HIGH ENERGY ELECTRONS AND CONTAMINATION ON LANGMUIR PROBE DIAGNO-STICS OF GAS DISCHARGED

doi: 10.11728/cjss1989.03.180 cstr: 32142.14.cjss1989.03.180
  • Received Date: 1988-10-22
  • Rev Recd Date: 1900-01-01
  • Publish Date: 1989-08-24
  • In the typical method of Langmuir Probe diagnostics of laboratory plasma, only iorr current correction is made. In fact, some electrons escaped from the filament, are accelerated through DCelectric field, they make electron temperature observed by probe "greater" than its real value. The error of electron temperature distorted by high energy electrons is as large as 80%-200%. At the same time, the measurement error of saturated electron current caused by high energy electrons is 70%-150%, that is also the measurement error of electron density. The error caused by the probe surface contamination is also discussed.

     

  • loading
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Article Metrics

    Article Views(1893) PDF Downloads(1378) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return